- 14 -
Gelovit 200
Description of the unit
Gelovit 200 is used to melt and store gel duplicating
material. The temperature of the duplicating material
can be constantly read off an electronic display.
z Lowerable operating temperature
Gelovit 200 is equipped with an electronic temperature
control. The automatic melting system starts after the
duplicating material is inserted at the push of button
: the material is melted and subsequently cooled
down to the operating temperature “T2”. By pressing
button , the temperature can be brought closer to
the reduced operating temperature “T3” within
approx. 40 minutes and contraction due to hardening
can thus be prevented. Because duplicating material
can be kept above the gelling temperature without
clumps occurring only for a short time, it is automati-
cally heated up to the operating temperature “T2”
again after 20 minutes.
z Provision time
The melting and cooling down of 6 kg of duplicating
material to the operating temperature takes approx. 4
hours. The strong mixer motor crushes even solid
cooled-down duplicating material, such as Castogel
®
and Wirodouble
®
.
z Built-in timer
By means of a timer, the unit can be set for the night
so that duplicating material is available at operating
temperature at the beginning of the next working day.
z Quality assurance
Gelovit 200 makes an important contribution to quality
assurance because the number of quality-reducing
melting operations can be monitored (see instructions
in section on “Operation”).
z Safety in the event of power failure
In the event of a temporary power failure, the
microprocessor registers the last operating mode and
continues the setting last selected. It is melted again
only when the duplication material has cooled down
below the „reduced operating temperature“.
z The temperatures are factory-set to 93°C for
melting, 42°C for processing and 39°C for working at
reduced operating temperature. The temperatures can
be varied and thus adapted to the duplicating material
used. For BEGO’s duplicating materials Castogel
®
and
Wirodouble
®
we recommend the factory-set
temperatures. The settings have to be changed for the
BEGO duplicating material Wirogel M (see „Check set
temperatures“ in section on Operation).
Fig. 1 1 main power switch, mains supply 4 cover
2 outlet 5 fuse
3 operating panel 6 documentation
Fig. 2 Function buttons
Button Press, briefly (= ) Press, long (= )
T 1 Melting, then automatic cooling to T2 Set T1
T 2 Cooling / heating to T2 Set T2
T 3 Cooling to T3 Set T3
+ + 1 (for settings) Set number of melting operations
until warning
- - 1 (for settings) Restart counting of melting operations
Set timer Display number of melting operations
thus far
Start timer