- 19 -
Gelovit 200
z Let duplicating material that has overflowed
harden and then collect it. Fill in with new
duplicating material at next melting operation.
z Clean housing with water and mild cleaning agent.
- Do not use cleaning agents in the duplicating
material container.
Fault Cause Remedy
No display The fuse of the unit Replace fuses on right side of unit.
after the unit is switched on. is defective. If fault occurs again, contact
Customer Service Department.
The mains socket is defective. Check socket.
No display, Display defective. Call Customer Service Department.
but double acoustic signal.
°F (or °C) flashes; The last set melting Duplicate as usual.
acoustic signal approx. operation takes place. Before switching off the unit,
30 seconds long. remove old duplicating material.
E 1 Temperature sensors Call Customer Service Department.
E 2 do not work.
E 3
E 4 Motor or rotary sensor Call Customer Service Department.
E 9 defective.
E 5 Clock defective. Call Customer Service Department.
E 6
E 8 after the unit is switched on Cover is not closed Close cover and turn
or interrupted acoustic signal. correctly. so that it is locked.
Care and maintenance
Elimination of faults