Installation
Check set temperatures
1. Switch on main power switch on right side of
unit: the display indicates the actual tempera-
ture in the duplicating material container.
2. Keep button „T1“ depressed until the acoustic
signal sounds twice: the display indicates the
set temperature desired for approx. 4 seconds.
3. If necessary, change the temperature with the
buttons "+" and "-".
4. Check/change T2 and T3 in the same way.
Notes
The factory-set temperatures are recommended for BEGO duplicating materials Castogel
®
and
Wirodouble
®
:
The temperatures for Wirogel M have to be changed to: T1 = 96 °C, T2 and T3 = 54 °C (Wirogel M
cannot be processed at reduced temperature)!
The factory setting of the temperature display is in °C. To change over to °F, call Customer Service.
Property damage!
The unit may only be installed on an
adequately stable surface.
The unit may only be used in an
enclosed space.
WARNING
Electrical hazard!
Prior to the electrical connection
check whether the specifications on
the rating plate correspond to the
power supply network. In the case
of uncertainty, consult an electri-
cian.
The machine is subject to the provi-
sions of class of protection I and
may only be connected to properly
grounded power sources.
NOTE
The new temperature is saved approx. 4 seconds
after the last input.
T1 melting temperature = 93 °C,
T2 operating temperature = 42 °C, (T2 cannot be set lower than T3.)
T3 reduced operating temperature = 39 °C. (T3 cannot be set higher than T2.)