11
5. TO SET THE PRESSURE OF EACH CHAMBER IN YOUR GARMENT
5.1 With the pump o, press the “Power” On/O buon.
5.2 The pump will display the hours of usage for 5 seconds. While the hour meter is being dis-
played (rst 5 seconds) Press and hold the “Up” and “Down” buon at the same me and
hold them down for 5 more seconds.
5.3 The next set of numbers will show the current pressure seng in the “LED Readout” and
the chamber associated with that pressure in the “Chamber Number LED”.
5.4 Change the pressure seng using the “Up” or “Down” buon unl you get to your desired
pressure.
5.5 Press the “Mode” buon to move to the next chamber and repeat the steps above.
5.6 Aer you have set the pressure in all 8 chambers, press the “Power On/O” buon and
your pump will start to operate.
6. PROGRAMMING YOUR PUMP TO OPERATE IN DIFFERENT MODES
The SC-3008-DL can run in a number of dierent modes, all of which are set in the following
manner:
6.1 With the Pump o, press the “Power” On/O buon.
6.2 The pump will display the hours of usage for 5 seconds. While the hour meter is being dis-
played (rst 5 seconds) Press and hold the “Mode” buon 5 more seconds.
6.3 Aer 5 seconds, you will see “1Hr.” in the “LED Readout”. This indicates that you are in the
Timed Shuto mode. In this mode, the pump will operate for a set period of me between
10 and 120 minutes, then automacally turn o. The pump comes preset to this mode and
1 hour so there is no reason to go through these steps if you wish to run your pump in the 1
Hour Mode. If you would like to adjust the amount of me for treatment in the “Timed
Mode”, simple press the up or down arrows while “1Hr” is showing to adjust the me.
Once your me seng is complete, press the “Power” On/O buon and your pump will
begin its cycle.
6.4 If you want to operate in a dierent mode, aer “1Hr.” shows press the “Mode” buon
again and you will see “COn” which indicates you are in connuous mode. In this mode,
the pump will not shut o unl you manually turn it o by pressing the “Power” On/O
buon. Once you see “COn”, if you want to proceed with this seng, press the “Power”
On/O buon and your pump will begin its cycle.
6.5 Your pump comes with a “Pre-Therapy” treatment opon. In this mode, the pump will run
for 10 minutes hing only chambers 6,7,8 over and over again. Aer 10 minutes, the pump
will automacally go into either the 1 Hour mode or the Connuous mode depending on
how you set up your pump. The SC-3008-DL comes preset with the Pre-Therapy mode
turned o. To turn it on, during the Mode setup, press the “Mode” buon again aer 6.4
and you will see “PrE” in the LED Readout. While this is displayed, press the “Up” buon to
change the number in the “Chamber Number LED” from 0 to 1. The “1” indicates that the
“Pre-Therapy” treatment has been acvated, a “0” indicates that it is o. To move from 1
to 0, simply press the “Down” arrow. Once this is set, hit the “MODE” buon again which
will return you to the 1 Hour Mode. Hit the “Power” buon to start your pump in this
mode, or the “MODE” buon again to move to the Connuous Mode. A Pre-Therapy treat-
ment can be used in either mode.
Please Note: The Pre-Therapy feature should not be turned on unl the pump has run for
at least 10 minutes on a paent. This will give the pump me to learn the proper pressure
sengs for each chamber.