3-6
Bio-Tek FL600
6HWWLQJWKH)LOWHU3RVLWLRQVLQ.&6RIWZDUH
The second and most important step in changing or adding filters to the
reader is changing the KC4 software filter settings to accurately reflect the
new filters and their position on the wheel. The values for filters in the
KC4 software must match the values of the filters in the reader to avoid
damaging the reader and invalidating data.
NOTE 1: Settings are stored on the PC, not the FL600. Remember to reprogram
the filter settings if you move the unit to another PC.