1. Press appropriate ISOTOPE SELECTION KEY
2. Press SETUP
3
. The calibration value for selected isotope is displayed on ETD (#1)
4. Use the ⇑⇓ keys to set new calibration value on ETD (#1)
5. Press ENTER
6. Press SETUP
 ++#&" .*'+'( $, *
1. Press OTHER KEY
2. Press SETUP
3. Use ⇑⇓ keys to set new calibration value on ETD (#1)
4. Press ENTER
5. Press SETUP
6. System returns to normal operating mode – the new calibration value is stored
*'+'(  !,$+$#)+#'&$, *
1. Press any standard ISOTOPE SELECTION KEY
2. Press SETUP
3. Current default value for selected isotope is displayed on ETD (#1).
Compare this value to “Appendix D” in Operation Manual
4. Use the ⇑⇓ keys to adjust isotope default value as needed on ETD (#1)
5. Press ENTER
6. Press SETUP
7. System returns to normal operating mode – the new calibration values are stored
The following method can also be used to determine the dial setting for a calibrated standard.
 .+#-#+. *,) % &+
1. Press OTHER KEY
2. ETD (#1) displays dial setting for OTHER key
3. Use the ⇑⇓ keys to adjust isotope default value as needed on ETD (#1)
4. Put source in well and take reading
5. Press another isotope key
6. Press OTHER and dial setting is back to default setting