EasyManua.ls Logo

Biolight Meditech S12 Vet - IBP Measurement

Biolight Meditech S12 Vet
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S12 Vet Monitor User’s Manual
13-2
13.3. IBP measurement
13.3.1. Monitoring Procedure
Please make an IBP measurement following below steps:
1. Plug one end of the IBP sensor cable into the monitor's IBP cable connector
and the other end link to the IBP sensor.
2. Refers to the IBP sensor manufacturer's instructions for exhausting air in the
IBP sensor, it make ensure no air bubbles in the sensor's entire tube.
3. Connecting IBP sensor to the patient, which makes sure the sensor and heart at
the same level.
4. Selecting correct pressure label based on the measured pressure. Specifically,
please refer to 13.5.2 Change the pressure label.
5. Refers to 13.3.2 IBP Sensor Zero for zeroing. During this process, the sensor
keeps stationary and the valve is open to the atmosphere.
CAUTION:
Before IBP measurements, it should make sure all IBP sensors are zeroed
properly.
Before IBP measurement, it makes sure no air bubbles in the IBP sensor which
result in erroneous pressure readings.
When intracranial pressure (ICP) measurements put on a sitting patient, the

Table of Contents

Related product manuals