S12 Vet Monitor User’s Manual
13-6
13.5.3. Setting up display types about Extended Pressure
If current pressure label is set to the extended pressure (P1 or P2), you could select
the type which displays in the parameter area, following these steps:
1. Select the parameter area or waveform area about the extended pressure for
entering the corresponding pressure menu.
2. SelectSetupsubmenu.
3. SetMeasurement:
All: Corresponding pressure in the parameter area shows all the pressure:
systolic pressure, diastolic pressure and mean pressure.
Mean only: Corresponding pressure in the parameter area only shows
the average pressure.
Auto: The system will automatically shows the pressure is displayed in
the parameter area or only the average pressure according to the
measured value of the extended pressure.
13.5.4. Setting the pressure sensitivity
The blood pressure value displayed on the monitor is average calculation about the
collected data over a period time. The higher the sensitivity, the faster the monitor
responds when the patient's blood pressure value changes, but the measurement accuracy
is lower. Inversely, the lower the sensitivity, the slower the response of the monitor when
the patient's blood pressure value changes, but the measurement accuracy is higher. When
monitoring critically ill patients, setting up a higher sensitivity is useful for timely
analysis.
You can set up the sensitivity of the current pressure, following these steps:
1. Select the IBP parameter area or waveform area for enter the corresponding
pressure menu.
2. SelectSetupsubmenu.
3. SetSensitivity.
13.5.5. Setting the IBP Waveform
You could set up the IBP waveform, following these steps: