S12 Vet Monitor User’s Manual
16-4
according to the relevant statement of the infusion set used. For example, the
Double-Dove tube declares that 20 drops of distilled water are equivalent to 1
mL±0.1 mL, so enter: 20 in the Drops/mL parameter setup.
16.3.1.5. Start DM measurement and adjust drip rate
Connect the infusion tube to the patient, start drip monitor (DM) measurement
through theStart/Stop button on the DM module, and adjust to the desired drip rate via
the flow clamp. The DM module indicator light switches from yellow to green and blinks
synchronously with the dropping of liquid drops.
16.3.2. Stop infusion
During the infusion or after the infusion is completed, press the “Start/Stop” button
of the DM module, and the indicator light of the module will switch to a yellow state. At
this time, the monitor will exit the DM function and will no longer perform drip monitor
measurement.
WARNING:
In the non-infusion drip monitor state, when infusion is completed, the monitor
will not stop the infusion and send out an infusion completion alarm. Just quit
DM function, but the infusion is still continuing. If the infusion needs to be
stopped, the liquid stop clamp of the tube set needs to be operated to stop the
infusion.
16.3.3. Infusion completion
When infusion is completed, the indicator light of DM module is switched to red
flashing state, the liquid stop clamp is automatically closed, block the pipeline, stop
infusion, and the monitor generates infusion completion alarm.
After recognizing the alarm, the medical staff confirms the alarm, separates the
infusion tube from the patient, presses the “Reset” button on the DM module, opens the
liquid stop clamp, takes out the infusion tube set, and finishes a drip monitor.