S12 Vet Monitor User’s Manual
16-6
16.6. Setting DM
16.6.1. Setting main unit
You can set the display of the DM main unit by the following steps as below:
1. Select the DM parameter area to enter 【DM】 menu.
2. Set 【Unit】of DM. The selected unit is displayed in the DM parameter area in
the form of a main unit.
16.6.2. Setting unit conversion parameters
In order to ensure the accuracy of the flow rate, you need to set the conversion
parameter between the number of drops and mL. The steps are as below:
1. Select the DM parameter area to enter 【DM】 menu.
2. Set 【Drop Per Milliliter】 of DM. The default value is 20.
16.7. DM check
The DM module has already been verified before leaving factory. Generally, the user
can directly measure it. Please contact the manufacturer or maintenance personnel
authorized by the hospital to repair the DM module when the following two situations
occur:
——The maintenance test is unqualified;
——Clinicians doubt the accuracy of readings.