S12 Vet Monitor User’s Manual
19-5
the recorder door;
2. Put the recording paper into the recorder with the thermal side which is
smoother up.
3. Close the door of the recorder, and pull some recording paper outside of the
paper out port.
4. Check the position of the recording paper to ensure that the recording paper is
aligned with the paper outlet.
CAUTION:
Must use the thermo-sensitive paper that meets requirements; otherwise, it will
lead to recording failure, bad-quality record or damage of thermo-sensitive
printing head.
Do not pull out the recording paper during recorder printing, otherwise the
recording meter may be damaged.
Unless for paper replacement or fault remedy, don’t keep the recorder door
open.
19.8. Clearing Jam Paper
While the sound of recorder operation or printing of recording meter is abnormal,
please first check whether there is paper jam in the recording meter. If so, please clear it
as per following steps:
1. Open the recorder door;
2. Pull out the recording paper, and cut off the wrinkle part;
3. Load recording paper once again and close the recording meter door.
19.9. Cleaning Recorder
After long-time service, some paper scrap and impurity will accumulate on the
printing head, and affect printing quality as well as the service life of printing head and
roll shaft. The recorder can be cleaned according to the following methods:
1. Before cleaning, the measures such as wearing anti-static wrist strap shall
be adopted to avoid the damage to recording meter resulting from static;