S12 Vet Monitor User’s Manual
7
Pulse width 100ms±10%
Rise and drop time ≤1ms
Alarm output
Indicates the inherent
delay in determining
the alarm status.
≤1s
Alarm delay time
from the monitor to
remote equipment
The alarm delay time from the monitor to remote equipment is ≤
2s, measured at the monitor signal output connector.
Alarm signal sound
pressure level range
Within a distance of one meter, the peak volume range of the
audible alarm generated by the equipment is 45 dB (A) ~ 85 dB
(A).
A.6 Data Storage
Trend data
Long trend: 1800h,minimum resolution is 10 min
Medium trend: 180h, minimum resolution is 1 min
Short trend: 6h, minimum resolution is 5 second.
Parameter alarm
event
At least 3000 parameter alarm events and associated parameter
waveform at the moment.
ARR events
3000 ARR events,and the parameter waveform related to the
time of event occurrence.
NIBP measurement
result
At least 2400 groups.
Holographic
waveform
At least 72 hours. The specific storage time depends on the
waveforms stored and the number of stored waveform.
A.7 Wireless network
Conforming
standards
IEEE802.11a/b/g/n
Operating
frequency
2.4GHz ~ 2.495 GHz, 5.15GHz ~ 5.35GHz, 5.47GHz ~