EasyManua.ls Logo

Biolight Meditech S12 Vet - Measurement Specifications

Biolight Meditech S12 Vet
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S12 Vet Monitor User’s Manual
8
5.725GHz, 5.725 GHz5.82GHz
Data security
WPA-PSKWPA2-PSK
Encryption
AESTKIP
A.8 Measurement Specifications
The product shall meet the following measurement specifications. If there is no
special indication, the definition of the index shall preferentially refer to the special
standard of the parameter.
A.8.1 ECG
A.8.1.1 Standard
Refer to standards of IEC 60601-2-27
Refer to standards of IEC 60601-2-25
A.8.1.2 Performance indicators
Electrosurgery
protection
Cut mode: 300W
Coagulate mode: 100W
Recovery time: 10s
In compliance with the requirements in clause 202.6.2.101 of
IEC 60601-2-27.
Lead-off detection
current
Measuring electrode: <100nA
Driving electrode (RL): <1 Ua
Tall T-wave
rejection capability
1.5Mv
HR averaging
method
Under normal circumstances, the 12 most recent RR intervals
are averaged to compute the HR.
If the last 3 consecutive RR intervals are greater than 1200ms
(i.e., HR is less than 50bpm), the 4 most recent RR intervals are

Table of Contents

Related product manuals