S12 Vet Monitor User’s Manual
49
Alarm messages Default priority Alarm messages Default priority
Tachy MED Missed Beat LOW
Brady MED A-Fib LOW
Nonsustained V-Tach MED A-Fib End LOW
Vent Rhythm MED ECG Noise LOW
PNC MED Irregular Rhythm LOW
PNP MED Irregular Rhythm End LOW
Pause MED
D.1.3 RESP Physiological Alarm
Alarm messages Default priority Cause and solution
RESP Apnea High
The patient is not breathing or the respiratory
signal is too weak to measure the respiratory
rate. Please check the patient’s condition,
check whether the electrode plate is placed
correctly and whether the connection of
electrode plate, cable and lead wire is firm.
RESP Artifact High
The patient’s heartbeat interferes with
breathing, thus making it impossible to
measure the breathing rate correctly. Please
check the patient’s condition and check the
connection of electrode plates, cables and lead
wires.
D.1.4 SpO2 Physiological Alarm
Alarm messages Default priority Cause and solution
SpO2 Search Pulse
Timeout
High
Can’t find a pulse for a long time. Please
immediately check the patient’s condition. If
the patient is normal condition, please
replace placement position of blood oxygen