S12 Vet Monitor User’s Manual
57
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
CO2 Sensor Off Med Completely clear
The CO
2
sensor is detached from
the patient or monitor.
CO2 Out of Range Low Not clearable
The measured data of CO2
module is out of range and needs
zero.
CO2 Zero Required Low Not clearable The sensor needs zero.
CO2 Sensor Over
Temp
Low Not clearable Check sensor.
CO2 Compensation
Not Set
Low Not clearable
The CO2 sensor was not
initialized. Set compensation and
initialize.
CO2 Sleep Mode Prompt /
The CO2 sensor is in sleep
mode. Please select the
measurement mode, CO2 can
enter the working state.
CO2 Check Sampling
Line
Low Not clearable
The CO2 sampling tube is
blocked or damaged; The
sampling tube is kinked or
compacted; The exhaust pipe is
blocked. Check the sampling
tube.
CO2 Check Adapter Low Not clearable Reinstall the airway adapter.
CO2 Zero In
Progress…
Prompt /
The CO2 module is being
zeroed.
CO2 Sensor Warm up Prompt / The CO2 module is warming up.
CO2 Self-Test Prompt / Module initialization
CO2 Sensor Faulty, E* Low /
Hardware or software errors,
contact after-sales personnel to
check maintenance.
CO2 Self-Test Error Low Not clearable Hardware errors, Replace sensor,