EasyManua.ls Logo

Biolight Meditech S12 Vet - Page 56

Biolight Meditech S12 Vet
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S12 Vet Monitor User’s Manual
6-7
Settings (Bed number) to selectAlarmsubmenu.
2. InAlarmsubmenu to check the current physiological and technical alarm
information of the main bed.
6.4.1.7. Reset Other Bed Alarm
In the Bed View Settings (Bed number)window→selectReset Remote
Alarmwhich in the Alarmsubmenu, the alarm of the corresponding remote monitor
(main bed) is reset. Only when this function is turned on can you reset other bed alarm.
For detailed, please refer to 7.12.1 Other Bed Alarm Reset.
6.4.1.8. Selecting Waveform
Other bed observation area can display up to 4 waveforms. Following these steps to
select the label name of the waveform you want to observe:
1. Select other bed observation interface area, in the pop-up windowBed View
Settings (Bed number)to selectWavesubmenu.
2. In proper order to select First Wave, Second Wave, Third Wave
and Fourth Wave, then select the label name of the waveform in the
pop-up list. If you select is Close, then the display of one waveform will
be turned off.
6.4.1.9. Selecting Parameters
Other bed observation can display all the online parameters. Select the parameter
label names you want to observe as follows:
1. Select other bed observation interface area, in the pop-up windowBed View
Settings (Bed number)
to selectParamsubmenu.
2. Open the parameter labels you want to observe in the displayed list of online
parameters.
WARNING
The data displayed will delay in other bed observation window. Don't rely on
other bed observation window for real-time data.

Table of Contents

Related product manuals