S12 Vet Monitor User’s Manual
7-3
current situation.
Prompt: Prompt patient and system status information.
7.3.3. Alarm Indicators
When an alarm occurs, the monitor will indicate it to the user through visual or
audible alarm indications:
Alarm signal
High priority
alarm
Medium priority
alarm
Low priority
alarm
Prompt Note
AlarmLamp
The lamp
quickly flashes
red with 1.4Hz
~2.8Hz, Duty
cycle 20%-60%.
The lamp slowly
flashes yellow
with 0.4Hz
0.8Hz Duty
cycle 20-60.
The lamp turns
cyan without
flashing, Duty
cycle 100.
/ /
Alarm Tone
Mode
ISO
DO-DO-DO
--
D
O-DO
-----
DO-DO-DO
--
DO-DO
-----
DO-DO-DO - DO- / /
Alarm Information
White words,
Red background
Black words,
Yellow
background
Black words,
Cyan
background
White
words
Displayed in the
top information
area, click on
the alarm
information to
view the alarm
information list.
Alarm level symbol *** ** * /
Displayed in
front of alarm
information.
Parameteralarm
Red
background,
flashes
Yellow
background,
flashes
Cyan
background,
flashes
/ /
WARNING
When multiple alarms of different priorities occur simultaneously, the alarm
lamp and alarm tone are prompted according to the highest level of all current