S12 Vet Monitor User’s Manual
7-11
7.8. Alarm Reset
Click onAlarm Resetquick key to reset the alarm system, and the alarm reset icon
will appear in the system status information area.
NOTE
In the alarm reset state, if a new alarm is generated, the alarm reset icon
disappears and the alarm system is reactivated.
7.8.1. Physiological Alarm Reset
After the physiological alarm is reset, the sound of the currently existing
physiological alarm is shielded, and the other alarm states remain unchanged.
7.8.2. Technical Alarm Reset
When the technical alarm is reset, it has the following characteristics:
The technical alarm that can be completely cleared is cleared. The monitor
will not have any alarm indication for the cleared technical alarm.
Technical alarm that can clear sound and light is displayed as prompt
message.
The sound of the technical alarm that cannot be cleared is shielded. For the
indication of the technical alarm after the alarm is reset, please refer to D.2
Technical Alarm.
7.9. Latching Alarms
The physiological alarms are classified into “Latching” and “Non-latching”.
Non-latching alarms:
After the condition that triggered the alarm of a
parameter disappears, the system will not make any prompt for this alarm of
this parameter.
Latching alarms: Even if the condition that caused the physiological alarm
disappears, the alarm signal will still be “Latched”, and the time of the last
triggering of the alarm will be displayed behind the alarm information in the