Reocor S Quick Reference Guide
Output (Ampl. V)
Up to 17 V
Two second tone if amplitude is programmed below 1.0 V
Programmable V Settings
Sensitivity (Sens. mV)
1.0-20 mV
Lights will indicate pacing or sensing
Basic Pacing Parameters
Mode Fixed or Demand
Rate 30-250 ppm
Two second tone if rate is programmed above 180 ppm
9 V Battery Drawer
Up to 600 hours of use
*
36 hours after battery warning
(Low Battery light blinks)
30 seconds during battery replacement
(more details on reverse side)
*
VVI, 70 ppm, 5 V, 500 Ω
Atrial Burst Functionality
60-1,000 ppm
Select the rate on the dial
When ready, press “Select burst”
followed by “Start burst”
Versatile Lead connection
Polarity configuration options
(more details on reverse side)
RF142r1 7/17/12
© 2012 BIOTRONIK, Inc. All rights reserved.