Display Reference
61
Specifications: BPM300
Pressure measurement
range
Pressure measurement
accuracy
Meets or exceeds ANSI/AAMI SP10:1992
Within 3 mmHg or 2%, whichever is
greater
Systolic blood pressure
measurement range
Diastolic blood pressure
measurement range
Pulse rate measurement
range
40 to 200 beats per minute
Blood pressure
measurement efficacy
Meets or exceeds ANSI/AAMI SP10:1992
standard for overall efficacy of blood
pressure measurements versus standard
auscultation (using the 5
th
Korotkoff sound)
Average error within +/- 5 mmHg
Standard deviation within 8 mmHg
Pulse rate measurement
efficacy
Meets or exceeds specification against a
clinical reference
Average error within +/-1 beats per minute
Standard deviation within 3 beats per
minute
Maximum operating cuff
pressure
Maximum possible cuff
pressure
Maximum possible cuff
inflation time
Over 10 mmHg for 3 minutes
Temperature probe sensor
type