BpTRU Operator’s Manual
66
Guidance and manufacturer’s declaration electromagnetic immunity
The BpTRU monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the BpTRU monitor should assure that it is used in
such an environment.
Immunity Test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 6100042
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
The BPM200 monitor should be
poweredoff or not taking a
measurement when the USB
cable is installed or removed.
Electrical fast
transients/ burst
IEC 6100044
±2 kV for
power supply
lines
1 kV for
input/output
lines
±2 kV for
power supply
lines
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 6100045
1 kV
differential
mode
2 kV
common
mode
1 kV
differential
mode
2 kV
common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000411
<5% UT for
0.5 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If you
require continued operation of
BPM100 during power mains
interruptions, power the
BPM100 from an
uninterruptible power supply
or battery.
40% UT for 5
cycles
70% UT for 25
cycles
<5% UT for 5
sec
40% UT for 5
cycles
70% UT for 25
cycles
<5% UT for 5
sec
NOTE: UT is the a.c. mains voltage prior to application of the test level.