BpTRU Operator’s Manual
70
Cables and acceptable lengths
Cable
Length
AC power cord-set
1.8 m (6’)
Oral (blue) temperature probe
1.8 m (6’)
Rectal (red) temperature probe
2.4 m (8’)
USB cable
3.0 m (10’)
CAUTION
The use of accessories, transducers, and cables other
than those specified may result in increased EMF
emissions or decreased immunity of the system to
EMF within the surrounding environment.
Technical Modes of Operation
Technical modes of operation are available for the
BpTRU to change configurations, check functionality and
accuracy, and to connect the monitor to a computer.
BP Alert/Silent Mode
As well as the power-up tone, the BpTRU makes a short
audible tone after completing a series of measurements
in Automatic mode. This feature is enabled by default,
but can be disabled and re-enabled as follows:
1. Turn the BpTRU off using the power switch.
2. To toggle BP Alert/Silent Mode, hold Stop while
turning the BpTRU on.
Repeating this toggles the BpTRU to the previous state.
Alert Mode is indicated by three short beeps followed by
one long tone during power up.
Silent Mode is indicated by three short beeps during
power up without a final long tone.