Welch
Allyn
Service Documentation
Part Number: 701627
Rev. G
Welch Allyn
701627 REV. G
Page 3 of 9
Special tools: The accuracy of infrared thermometers can only be tested with a blackbody device
(BB-3200 Series, no longer available for purchase) or the 9600 PLUS Calibration
Tester.
The Model 9600 PLUS Calibration Tester can be ordered through Welch Allyn
Customer Service (see page 8) for Welch Allyn contact information.
Model Welch Allyn P/N Braun P/N
9600 PLUS Calibration Tester 01802-110 N/A
The following additional data will be necessary:
Bill to address
Ship to address
Quantity
Billing contact person with fon/fax extension
Shipping contact person with fon/fax extension
Shipping method
Purchase Order / Credit Card
The blackbody has to be calibrated every 12 months by the manufacturer. Contact
Welch Allyn Customer Service to get a Return Material Authorization number to have
your unit calibrated. Refer to Customer Service Listing (see page 8) for Welch Allyn
contact information.
Please direct all requirements and questions concerning blackbodies directly to the
Welch Allyn Service Center. Refer to Customer Service Listing (see page 8) for
Welch Allyn contact information.