Accessories functional check
Accessories Passed Failed
Test resistor 10 kΩ +/- 1 %
□□
Electrode cable
□□
Stimuplex® Remote Control, if available
□□
Stimuplex® Pen, if available
□□
Stimulation signal check
Test Stimuplex® HNS 12 Setting Tolerance Passed Failed
Pulseform 0.10 ms, 1.00 mA Square, negative
□□
Stimulation duration 0.10 ms, 1.00 mA 0.098 ms – 0.102 ms
□□
Stimulation frequency 2 Hz, 1.00 mA 1.98 Hz – 2.02 Hz
□□
Stimulation current 0.10 ms, 0.30 mA, 10 kΩ +/- 1% 0.26 mA – 0.34 mA
□□
Stimulation current 0.10 ms, 0.50 mA, 10 kΩ +/- 1% 0.46 mA – 0.54 mA
□□
Stimulation current 0.10 ms, 1.00 mA, 10 kΩ +/- 1% 0.96 mA – 1.04 mA
□□
Stimulation current 0.10 ms, 5.00 mA, 10 kΩ +/- 1% 4.96 mA – 5.04 mA
□□
Impedance 0.10 ms, 5.00 mA, 10 kΩ +/- 1% 9 kΩ – 11 kΩ
□□
Remarks:
Technical inspector: Date: Signature: