12
Error Description Potential cause and remedy
Batteries are low Insert new batteries
Batteries are dead Insert new batteries
«ERR 1» Signal too weak The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the
measurement.*
«ERR 2» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping the arm still.
«ERR 3» No pressure in the cuff An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is
not too loose. Replace the batterries
if necessary. Repeat the measurement.
«ERR 5» Abnormal result The measuring signals are inaccurate and
no result can therefore be displayed. Read
through the checklist for performing
reliable measurements and then repeat the
measurement.*