How to Use the Clearway 2 42
DOC-001000 Ver 6 Clearway 2 User Manual .
Preview Treatment
When the Clearway 2 is in standby and
in any of the automatic MI-E modes,
including a saved TreatRepeat
programme, or NIV mode the
programmed treatment waveforms
can be previewed by pressing the
Preview Treatment button. Pressing
the arrow buttons moves the view forward or backward through the treatment
programme. In Manual Mode it is only possible to View Treatment that is
being delivered by the manual switches.
It is not possible to Preview Treatment in the IPPB mode
Active Treatment Display
When the Clearway 2 is delivering treatment in an automatic Mi-E mode, you
can press the Preview Treatment button to see a visual representation of the
programmed treatment.
A vertical line follows the
treatment as it is being
delivered. This may help
patients to synchronise with
the Clearway 2 when using it
by allowing them to keep
track of the treatment
programme.
In the Manual Mode, users can press the View Treatment button to see live
feedback waveforms of the pressure being delivered.