Alarms 50
DOC-001000 Ver 6 Clearway 2 User Manual .
Alarm Reset
An alarm condition will be automatically reset when the cause of the alarm has
been corrected or the alarm condition no longer exists.
If an alarm cannot be corrected, stop using the Clearway 2 and
request technical support
When an alarm has been acknowledged on screen, the alarm message will
disappear unless the fault is reoccurring or still exists. If the fault still exists
after acknowledgment, the alarm symbol will remain on screen and enables the
user to view the current list of live alarms.
Alarm and Warning descriptions
High Pressure Alarm
120% of set pressure is detected for > 2 seconds
Treatment stops. The on-screen message must be
acknowledged. Treatment cannot be started until
message has been acknowledged
Treatment has stopped!
- Restart the Clearway 2
- Contact your supplier if fault persists”