69 Technical Specifications
Clearway 2 User Manual
DOC-001000 Ver 6
Environmental Operating Conditions
SPECIFICATIONS
Range
5 to +40°C (41 to 104°F)
Storage and Transport
Temperature
-20 to +60°C (-4 to +140°F)
Applied Parts Maximum
Operating Temperature
+43°C (109.4°F)
Ambient Pressure Range 700 hPa to 1,060 hPa
Humidity 15% to 90%, non-condensing
OPERATING CONDITIONS SPECIFICATION
Sound level at 70 cmH
2
O
in Manual Mode
Less than 74 dB(A)
Measured at 1 m
EXPLAINATION
IP22
Solid particle protection: Hazardous parts are protected
from touch by fingers and by objects greater than 12 mm.
Liquid ingress protection: The protection withstands
dripping water less than 15 degrees from vertical.
The ingress protection has been tested by water drips
equivalent to 3mm rain/minute for 10 minutes (2.5
minutes for each tilting direction)