9Safety Information
iSleep 20 clinician’s manual
Doc. 003923 En-Us V-2
2.3 Environmental Conditions
Do not use the iSleep 20 in any toxic environment.
Do not use the iSleep 20 in environments where there are explosive gases
or other flammable anesthetic agents present.
The performance of the iSleep 20 may deteriorate at:
ambient temperatures below 41°F (5°C) and above 104°F (40°C).
ambient relative humidity below 10% RH (relative humidity) and above
95% RH.
atmosphere pressure below 700 mbar and above 1060 mbar.
Do not use the iSleep 20 while positioned in a warm place, such as direct
sunlight.
The device complies with the EMC requirements of standards listed in
“Compliance of Standards” on page 49. Necessary measures should be
taken in order to assure field levels exceeding 10 V/m are avoided, since
this may impair the safety and performance of the iSleep 20. Measures
should include but not be limited to:
normal precautions with regard to relative humidity and conductive char-
acteristics of clothing in order to minimize the build-up of electrostatic
charges.
avoiding use of radio emitting devices closer than 1 m to the iSleep 20.
Radio emitting devices are for example cellular or cordless telephones,
microwave ovens and high-frequency surgery apparatus.
The iSleep 20, any accessories and all replaced parts must be disposed of in
accordance with the local environmental regulations regarding the disposal
of used equipment and waste.
The performance of the iSleep 20 and treatment of the patient may deteri-
orate if the operation conditions in “Technical Specifications” on page 46
are not fulfilled. Do not use the iSleep 20 immediately after storage or
transport outside the recommended operating conditions.