AMETEK Brookeld Page 12 Manual No. M09-1200-F1016
I.7 Key Functions
Bubble Level
Access Door.
Lift ap to view
Bubble Level
Test Stop
Button
POWER
Figure I-3: Control keys for standalone mode
The control keys are used only for manual positioning of the compression lid assembly and the
sample trough drive. Powder Flow Pro™Softwareisrequiredforallotheroperations.
Keypad
This key is used to move the compression plate assembly UP away from the sample trough
drive assembly. It is also commonly used for general purposes.
This key is used to move the compression plate assembly DOWN towards the sample trough
drive assembly.
PThis key is used to rotate the sample trough drive assembly clockwise.
Note: This key is also used during calibration.
QThis key is used to rotate the sample trough drive assembly counter-clockwise.
Note: This key is also used during calibration.
I.8 Test Stop - Reset
This button is used in case of an emergency situation. When this button is depressed, all
software control is blocked and movement halts. Note that the manual control buttons on the
instrument can still be used to reposition the lid and trough as necessary. Once the emergency
situation is cleared, this button must be rotated to reset. After using the Test Stop button, both
the instrument and the software should be restarted to ensure that normal communications have
been reestablished.