System Maintenance
Global User Manual
9-5
System Maintenance
Note: The scheduled intervals are based on an 8-hour day and a
40-hour week.
If the machine is operated greater than the scheduled interval
(e.g. two 8-hour shifts a day vs. one), adjust the particular
interval as necessary, and continue to perform the preventative
maintenance at that scheduled interval (before the first 8-hour
shift and before the second 8-hour shift).
If the machine is operated less than the designated interval (4
hours a day), continue to perform the preventative maintenance
on a daily (once a day), a monthly (once a month) and a quar-
terly (once every three months) basis.
Scheduled Maintenance Checklist
Daily:
ü Inspect the CMM for any loose or damaged covers. If
necessary, secure any loose covers and repair any damaged
covers.
ü Wipe all the exposed surfaces of air bearing travel ways with
alcohol and a clean, soft cloth towel (medical gauze, or
equivalent).
ü Verify system air pressure indicates 4.0 BAR. Adjust system
air pressure, if necessary. Refer to the Adjustments &
Alignments Section.
Caution:
Do not operate the CMM until the checklist is
completed and all the discrepancies fixed.