EasyManua.ls Logo

BRUEL & KJAER 1024 - 3 kHz Filter Verification; 3 kHz Oscillator Frequency; Noise Generator Calibration

BRUEL & KJAER 1024
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.3
. 3 k
Hz
Filt
",
o.
MATCHING IMP
.:
"600"
COMPRESSOR
SP
EED:
"Comp.
Ofr'
SW
ITCH
07:
"Scanning Off"
BANDWID
TH
SELECTOR:
"10Hz"
METER
SWITCH:
110.3
sec .
Output
Vo
lt
,"
b.
Check
BANDWIDTH
SE
LECT
OR
in
all
positions
BANDWIDTH SELECTOR:
Before adjustment
of
the coils toke the instrument out
of
its
case
and
leave it
swi tched
on
for
2 hours
min
.
This
ensures that condensers and
coils
hove the
same temperature .
Remove
tube
V 13
and
connect
ty~
1022
to
filt
~
r
input
(R
115).
Input
signal: 2
mV
3000
Hz
t
0.3
Hz
.
Adiust Z 3
and
Z 4 for max .
defleclion
on type
l024.
Change
the
frequency appro
x.
+ and - 100
Hz
and check the top
of
the filter
curve.
Jt
should be flat
wit
hin 2
dB
and the height
of
the three tops should be
within -
0.5
dB
.
10
Hz
30
Hz
1
00
Hz 300
Hz
Frequency
Damping Frequency
Hz
dB
Hz
3ooo!
0
0
3000::
5
3
10
min.
10
20 20
30
30
3.4.
3
kH
z
Oscillator
BANDW
IDTH
SELECTOR: "Sine"
3.5.
Noise Generator
MATCHING
IMP
.:
"600
0"
COMPRESSOR
SPEED:
"Comp.
Off"
SWIT
CH
07:
"Sc:onning OWl
BAND
WI
DTH
S
ELECT
OR:
"Sine"
METER
SWIT
CH
: "
10
sec.
Output
Volt. "
FREQUENCY SCALE:
"1000
Hz"
b.
BANDWIDTH
SELECTOR
to
20-20000
Hz
c.
B
AN
D
WIDT
H
SE
LECTO
R in pos
it
ions
10 -
30-100-300Hz
d.
BA
NDWIDTH
SE
LECT
OR to
20-20000
Hz
1024.3
-
2.68
0
15
30
60
90
Damping Frequency Damping Frequency
Damping
dB
Hz
dB
Hz
dB
3000 ±
0
3000+
0
0
50
-150
3
min . 20
1
00
min .
20
300
min.
20
40
200
40
600
50
50 300
50
Short-circuit 91 on V 1 (pin 2)
to
ground and
connec
t a frequency counter an
the
sec9f\dary
af
Z
4.
The
frequency should be: 3000 H
z.
Tolerance::
1 Hz.
Ad
j
ust
P 8 to 200
mV
measured an the secondary
of
Z
4.
Connect an
oscilloscope
to LOAD terminals and adjust the
OUTPUT
LE
V
EL
for
a 20
dB
deflection
on
type
1024.
Check that
th
e amplitude distribution
of
the noise
is
symmetrical .
If
necessary adjust P
2.
Check
the
defl
ect
ion on
type
1024:
19
- 21
dB
.
If
necessary adjust P 7 .
Check
the
deflection
on
type
1024
:
19
-
21
dB.
If
necessary adjust P
3.

Related product manuals