3.
OPERATION
3.1.
PRELIMINARY
ADJUSTMENTS
Before
connecting
a
mains
supply
the
following
checks and
adjustments
should
be
car-
ried
out
to
ensure
safe operation
of
the
apparatus.
3 .
1.1.
Selection
of
Line
Voltage
The
instrument
can be operated on
single
phase AC
line
voltage
from
100
V
to
240
V,
with
a
tolerance
of
±
10%
,
at
frequencies
between
50
and
400
Hz.
Set
the
LINE VOLTAGE SELECTOR
to
the
voltage
to
be used by
turning
it
with
a
small
coin
or
a broad-bladed
screwdriver
.
3.1.2.
Fuse
Rating
For
operation
on
100,
115
and
127
V,
two
0,2
A
fuses
(VF
0012)
are
required, one in
each
holder
.
Operation
with
200,
220
and
240
V
requires
only
one
0,2
A fuse, and
when
these
voltages
are
selected as described
in
3 .1.
1,
the
left-hand
fuse is
electrically
by-passed.
Make
sure
that
only
fuses
with
the
required rated
current
and
of
the
specified type are
used
for
replacement.
The use
of
mended
fuses
and
the
short-circuiting
of
fuse
holders
shall
be avoided.
3.1.3
.
Line
Connections
Blue
Neut ral
AC Power
'"
B
row
n
Suppl y
Li
ve
[
I~
I
J
Gr
ee
n and Yell
ow
Gr
ound
172246
Fig.3 . T.