insulated,
it
supplies 1 2 V
DC
at
up
to
400
rnA
.
Colour
coding
for
the
mains
power
cable
is as
follows:
Blue -
Neutral
Black -
100
to
130V
AC
Brown
-
200
to
240V
AC
Fig. 3. 7.
Battery
Charger
ZG
0113
The
ZG
0113
is
terminated
with
a
7-pin
DIN
plug
which
plugs
into
the
EXT. POWER
SUPPLY socket
of
the
1623.
Expected
life
for
the
fully
charged Nickel-
Cadmium
battery
is
about
8 hours.
Alternative
methods
for
charging
the
battery
are
mentioned
under
sec-
tions
3.2.3
and 3 .
2.4
.
WARNING:
There
is a
danger
of
explosion
if
recharging
is
attempted
on
cells
that
have
been
fitted
with
their
polarities reversed.
Time
taken
for
a
full
recharge
of
the
Ni-Cd
cells
is
about
14
hours.
If
the
cells
are
not
fully
charged
after
this
period,
replacement
is
indicated.
A
minimum
of
500
recharge
cy-
cles
is specified
for
these
cells
but
their
lifetime
can be extended
if
they
are
not
allowed
to
become
fully
discharged too
often.
Never
attempt
to
recharge
any
type
of
non-rechargeable
cells
-
there
is
a
danger
of
explosion.
Whenever
a
charging
voltage is
applied
to
the
EXT. POWER SUPPLY socket (Pin 3),
charging
power
is being applied
to
the
cells
in
the
1623.
The
position
of
the
slide
switch
marked
Ext.
Power-
Int.
Battery
on
the
rear
panel,
is
irrelevant
.
Note
that
the
Battery
Charger
Unit
ZG
011
3
has
a
current
capability
which
is
approxi-
mately
equal
to
the
current
consumption
of
the
1
623
when
operating,
so
that
with
well-
charged
cells,
the
ZG
0113
will
act
as
an
AC
Mains
power
adaptor.
With
this
method
of
powering
/
charging
the
1623
the
rear
panel
power
supply
slide
switch
should
be
switched
to
Int.
Battery.
The
output
of
the
ZG
0113
is
short
circuit
protected by a
thermal
overload
cut-out
which
breaks
the
mains
power
supply
when
the
primary
windings
reach a
temperature
of
ap-
prox
.
80°C
.
Power
is reconnected again
automatically
after
a period
of
10
to
30
mins.
depending
on
the
ambient
temperature
.
3.4.
FUSE
REPLACEMENT
16
The
instrument
is protected by a
0,63
A
slow-blow
fuse
B & K Type VF
0032.
Access
to
the
fuse
is
gained
by removing
the
two
screws
securing
the
top
panel
and
sliding
the
panel
backwards
. The
fuse
holder
is
positioned
in
the
upper
right-hand
corner
(viewed
from
the
front)
of
the
cabinet.