EasyManua.ls Logo

BRUEL & KJAER 2209 - Specifications; Frequency Response

BRUEL & KJAER 2209
119 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.
SPECIFICATIONS
Specifications
for
sound measurements refer
to
the meter
with
input
stage,
flexible extension rod
and
1 inch microphone 4145 unless otherwise
stated.
8.1. FREQUENCY RESPONSE
(see
also Fig.4.9
for
low
frequency
responses)
One
inch microphone Type 4145
Linear
Sensitivity
Temperature Coefficient:
Temperature
Range:
Long Term Stability:
Half inch microphone Type 4133
Linear
Sensitivity:
Temperature Coefficient:
6-
4.000
Hz
4-
12.000
Hz
3-
18.000
Hz
±0.5
dB
± 1 dB
+
1
dB
-1.5
approx. 50
mV
per
Nfm2
0.006
dB/°C
-50
to
150°C
± 0.1 dB
plus<+
0.1
dB per 10 years at
operating temperature below
35°C.
At
100°C
typically+
0.2 dB per 100 hours.
10-
8.000 Hz
6-20.000
Hz
4-40.000
Hz
±0.5
dB
± 1 dB
±2
dB
approx.
10
mV
per
Nfm2
<±0.006
dB/°C
Individual calibration charts
are
supplied
with
every microphone.
106