Maximum Output Voltage: 100 V peak
to
peak.
Noise:
At
least
50
dB
below FSD.
DC
OUTPUT '
Output
Impedance:
25
kU.
Direct short circuit
of
the
DC
OUTPUT will
not
affect the meter reading.
Output
Voltage corresponding
to
FSD
(open circuit): Approx. 1 V RMS.
8.7. METER DAMPING
"Fast"
and
"Slow"
in accordance
with
I
EC
Recommendation 179.
"Impulse"
in accordance
with
proposed extension
to
I
EC
179.
"Imp.
Hold"
indicates maximum
RMS
level
of
input
signal. Integrating
time constant 35
msec.
Decay time < 0.05 dB/sec.
"Peak
Hold"
indicates maximum PEAK Level
of
input
signal.
Rise
time
<20
JJ.Sec.
Decay time
<0.05
dB/sec.
8.8. METER
INDICATION
RMS (Meter Switch in position
"Imp.
Hold",
"Imp.",
"Fast",
"Slow"):
Crest Factor Capability:
Meter Accuracy:
10 at
FSD
increasing
to
40 at
12
dB
below
FSD.
For crest factor
up
to
10.
±0.5
dB
from
FSD
to
12
dB
below FSD.
± 1
dB
from
12
dB
to
20
dB
below FSD.
These
limits
are
increased by 0.5
dB
for
crest factors between 10
and
20,
and
by 1 dB
for
crest factors between 20
and
40.
See
also Fig.4.15 and
Table 4.2.