BRUEL & KJAER 2221 - Leq and SEL Measurements; Max Hold Fast;Slow Measurements

BRUEL & KJAER 2221
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
For
measurements
in a
diffuse
field,
the
Random
Incidence
Adaptor
DZ 9566
should
be
mounted
on
the
microphone,
and the
meter
may
be
oriented
in
any
direction
in
the
diffuse
field.
A
diffuse
field can
generally
be
assumed
for
indoor
measurements
or
in
reverbera-
tion
rooms.
3.5.3.
Leq
and
SEL
Measurements
1.
Carry
out
the
calibration
check
before
commencing
with
a
measurement:
2.
Set
the
controls
on the
Sound
Level
Meter
as
follows:
FUNCTION:
Leq
or
SEL
Leq
RANGE: in
order
to
include
the
expected
value
MODE:
"Run"
POWER:
"On"
3.
To
begin
measurement,
press
RESET.
The
Leq
or
SEL value
measured
can
be
read
directly
from
the
display
which
is
updated
twice
per
second
while the
measurement
is running.
Alternatively
the
measurement
can
be
interrupted
by
setting
the
MODE
switch
to
"Pause".
The
"A"
displayed
indicates
that
A-weighting
is applied.
If
an
input
overload
occurs,
the
sign
"•"
appears
(held
continuously)
and the
reading
may
be
incorrect.
Switch
Leq
RANGE
to
a
higher
range.
If
the
input
signal
is
underrange
the
sign
"'f'"
appears.
If
it
is
displayed
too
often,
switch
Leq
RANGE
to
a
lower
position.
Switching
between
measuring
ranges
automatically
resets
the
instrument.
The
Leq
measurement
duration
for
periods
shorter
than 2,77
hours
can
be
obtained
by
substracting
the
measured
Leq
from
the
SEL value and using
the
conversion
table
at
the
back
of
the
instrument.
SEL is
obtained
by
setting
FUNCTION
to
"SEL".
See Fig.3.7.
The
Leq
value
displayed
for
measurements
over
a
period
of
more
than 2,77
hours
will
be
incorrect
and a
"*"
will
appear
on
the
display.
However,
knowing
the
measurement
duration,
the
correct
Leq
value can
be
obtained
from
the SEL
indication
using
the
conversion
table
at
the
back
of
the
instrument.
The
measurement
period
for
an SEL
measurement
is
only
limited
by
the
battery
life.
Note:
It
is
possible
to
switch
between
"Leq"
and
"SEL"
during
a
measurement
without
loosing
the
results. When
measuring
low
levels
it
is
recommended
to
set
the
MODE
switch
to
"Pause"
before
switching
between
"Leq"
and
"SEL",
as the
clicking
noise
from
the
switch
may
introduce
an
error.
This is
irrelevant
when
measuring
higher
levels.
3.5.4.
Max
Hold
"Fast"
or
"Slow"
Measurements
14
1.
Carry
out
the
calibration
check
before
commencing
with
a
measurement.
2.
Set
the
controls
on the
Sound
Level
Meter
as
follows:
FUNCTION:
Max
Hold
"Fast"
or
"Slow"
Leq
RANGE:
MODE:
POWER:
to
include
the
expected
value
"Run"
"On"

Related product manuals