BB106411.book : 2237-ch3 Black 10
Chapter 3 – Taking Measurements
Taking Measurements
3–10
Brüel & Kjær2237 Controller
User Manual
If an overload has occurred, then stop measuring (if neces-
sary), increase the measurement range by pressing and
retake the measurement.
Overloads are also indicated in printouts and in records.
● In recalled records, all parameter values will blink if an
overload occurred in that record (see section 4.1.3)
● Printed results indicate an overload by showing “Over-
load
” on the printout (see section 4.2.2)
● Spreadsheet results have a special overload column in
which overloads are indicated with “Y”.
Fig.3.5 Latched overload
RUN 00:05:41
MaxP 94.7 dB
AF Leq 45.8 dB
30 100
Upper limit of the measurement
range (overload if blinking)