Chapter 4 – Data Operations
Transferring Records to a Computer
4–16
Brüel & Kjær2237 Controller
User Manual
4.3.3 Transferring Results with Third Party
Software
In some situations (e.g. if you are using a Macintosh™ or
DOS-only machine), you may want to use communications
software other than that included with your instrument. If
so, follow the instructions below.
1. Connect the 2237 Controller to your computer as de-
scribed in section 4.3.1.
2. Start your communications software.
3. Set your software’s communication parameters as follows:
● 9600 Baud
● 8 data bits
● 1 stop bit
● Parity: none
● Handshake: XON/XOFF
4. Your communications software will probably ask you for
a name and location to which to save the data. Make the
appropriate selections. All records will be saved in a sin-
gle spreadsheet format file (tab separated) with the name
and location you choose.
5. On the sound level meter, press until the Output
Stored Results Window is visible on the screen, as shown
below.
Fig.4.16 The Output Stored Results Window
OUTPUT
Stored Results
to
Spreadsheet