EasyManua.ls Logo

BRUEL & KJAER 2240 - Indicator Range; Linearity Range; Pulse Range; Linear Operating Range

BRUEL & KJAER 2240
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 4
Specifications
19
4.9.4 Indicator Range
4.9.5 Linearity Range
Linearity Range according to the International Standard IEC 60804 is the difference between the Upper
and Lower Limit in the following table:
4.9.6 Pulse Range
Pulse Range according to International Standard IEC 60804 is the difference between the Upper and
Lower Limit in the following table:
4.9.7 Linear Operating Range
4.9.8 Peak C Range
Table 4.6
Indicator Range
according to Inter-
national Standard
IEC 60804
Range Name
Upper Limit
(dB)
Lower Limit
A-weighting (dB) C-weighting (dB)
1 140 60 60
2 110 30 36
Table 4.7
Linearity Range
according to Inter-
national Standard
IEC 60804
Range Name
Upper Limit
(dB)
Lower Limit
A-weighting (dB) C-weighting (dB)
1 140 60 60
2 110 30 36
Table 4.8
Pulse Range ac-
cording to Interna-
tional Standard
IEC 60804
Range Name
Upper Limit
(dB)
Lower Limit
A-weighting (dB) C-weighting (dB)
1 143 60 60
2 113 30 36
Table 4.9
A-weighted
Linear Operating
Range
A-weighting 31.5 Hz 1 kHz 4 kHz 8 kHz 12.5 kHz
110 – 30 dB
Upper Limit 70.0 110.0 111.0 108.0 105.0
Starting Point 54.0 94.0 94.0 94.0 94.0
Lower Limit 30.0 30.0 30.0 30.0 30.0
140 – 60 dB
Upper Limit 100.0 140.0 141.0 138.0 135.0
Starting Point 84.0 124.0 124.0 124.0 124.0
Lower Limit 60.0 60.0 60.0 60.0 60.0
Table 4.10
Peak C Range
according to Inter-
national Standard
IEC 61672 1
dB
110 – 30 dB
Upper Limit
113.0
Lower Limit
60.0
140 – 60 dB
Upper Limit
143.0
Lower Limit
80.0

Other manuals for BRUEL & KJAER 2240