Measuring Ampifier Type 2525
User Manual Vol.2
Brüel & Kjær
6–14
Chapter 6 Set-up and Display (Results) Messages
Measurement Set-up Commands
6.3.9 Peak_Hold_Time
The syntax for this command is shown in Fig.6.22.
The allowable Times are:
Time (Range): 0.5 to 60.0 s
Resolution: 0.5 s
Round: Nearest 0.5 s
Default: 5 s
The syntax diagram for the query Peak_Hold_Time? and the amplifier response is
shown in Fig.6.23.
Fig.6.22 The syntax for the command Peak_Hold_Time
Fig.6.23 The syntax for the query Peak_Hold_Time? and amplifier response