Measuring Ampifier Type 2525
User Manual Vol.2
Brüel & Kjær
6–22
Chapter 6 Set-up and Display (Results) Messages
Display Set-up Commands
6.4 Display Set-up Commands
The syntax for the Display_Set_Up command appears as shown in Fig.6.36.
The default value is:
CH1_Rms, CH1_POsitive_Peak, CH1_Negative_Peak
Note: With the exception of OFf, you must select three different parameters for the
Display_Set_Up command line. Thus:
Display_Set_Up CH1_Rms,CH1_POsitive_Peak,CH1_Rms
is illegal and results in device error 103.
However:
Display_Set_Up CH1_Rms,CH1_POsitive_Peak,CH1_PEak_Peak
is legal, as is:
Display_Set_Up, CH1_Rms,OFf,OFf
The syntax for the query appears as shown in Fig.6.37.
Fig.6.36 The syntax for the command Display_Set_Up
941158e
DISPLAY_SET_UP
To 2525
:
Te
Display par
Display par
Char
SP
,
,
Display par
Char
see below
see below
see below
CH1_PEAK_PEAK
CH1_NEGATIVE_PEAK
CH1_POSITIVE_PEAK
CH1_RMS
OFF
OFF