Loading...
the 5 Amps
RMS
current
limit.
It
is
not
recommended
to
use
loads below
2 n
as
power dissipation on the heat sink
will
rise considerably.
When
the 1.8 Amps
RMS
current
limit
is
selected,
it
does
not
affect the
voltage
limit
which remains 15
Volts
RMS. Maximum power
will
therefore
be
available
with
a load
of
about 8
n.
The 1.8 Amps RMS
limit
can
be
used
to
drive the Mini-Shaker Type 4810 safely
at
full
rating. Fig.4.3 shows the
maximum power
available
as
a
function
of
load resistance
for
the
two
current limits.
75
~
[
"
50
a.
"
0
x
~
25
2 3 4 5 6 7 8 9
10
11
12
13
14
15
r.nool
Load Resistance
In
I
Fig.4.3. Maximum
Output
Power
for
both
current
limits
A typical frequency
response
curve
is
shown in Fig.4.4. The specified
frequency
range
of
Type 2706
is
10
Hz
to
20 kHz
0.5 dB).
1
(
\
\
-2
I \
I
i\
-3
1
10 20 50 100 200
500
1 K
2K
SK
lOK
20K
SDK
lOOK
Frequency (Hz)
¥l.Z
""
~
Fig.4.4.
Amplifier
Frequency Response
(10
ARMS
into
a 2 n load)
11