Chapter 2 — Prepolarized Free-field
1
/
2
" Microphone Type 4188
Introduction
Falcon™ Range of Microphone Products
Microphone Handbook
BE1374– 12
2
−
3
2.1.2 The Calibration Chart
Each microphone is supplied with an individual calibration chart (see Fig.2.2)
which gives the microphone’s open-circuit pressure sensitivity together with the
typical capacitance and free-field and random-incidence frequency responses
*
. When
these are combined with the microphone’s typical data supplied in this chapter, the
individual microphone’s response under various conditions can be determined.
Open-circuit Sensitivity
The stated open-circuit pressure sensitivity is valid at the reference frequency
(1000Hz) for random-incidence and pressure-field conditions. The free-field sensitiv-
ity at the reference frequency (1000Hz) is 0.11dB higher than the pressure sensi-
tivity.
Ambient Conditions
The ambient conditions are measured continuously during calibration at the factory.
The calibration results obtained at the measured environmental calibration condi-
tions are corrected to the stated reference ambient conditions (23°C, 101.325kPa
and 50%RH).
Frequency Responses
Two typical frequency responses are shown on the calibration chart. Both are nor-
malized to 0dB at the reference frequency (1000Hz).
The left-hand curve on the rear side of the calibration chart is the open-circuit 0°-
incidence free-field response for the microphone without the supplied Random Inci-
dence Corrector DZ9566.
*Random-incidence response with supplied Random Incidence Corrector DZ9566.
Fig.2.2 Microphone calibration chart (front and back)
Brüel & Kjær
Caution: Static electricity discharge directly on the centre terminal may damage the prepolarization of the
cartridge. Therefore, ensure that the housing of the cartridge makes contact before the centre terminal.
Sensitivity: The loaded sensitivity is typically 0.05dB lower than the sensitivity stated. The random-field
sensitivity is the same as the pressure sensitivity. The free-field sensitivity at 1000Hz is 0.15dB higher
than the pressure sensitivity.
Free-field calibration with Sound Level Calibrators at 1000Hz: Adjust the Sound Level Meter, or other
measurement equipment, to indicate 0.15dB lower SPL than the actual SPL produced by the calibrator.
The two grooves means “prepolarized”, i.e. 0V external polarization voltage.
Refer to the 4188 Product Data for further information. See also rear side.
Prepolarized
Condenser
Microphone
Cartridge
Type
4188
Serial No.:
Open-circuit Pressure Sensitivity at 1013 hPa 23˚C and 50% RH:
dB re 1 V/Pa or mV/Pa
Frequency: 1000Hz Capacitance: 12pF (typical)
Polarization Voltage (external): 0V
Date: Signature: .......................................
B
K
7/6-'89
BC0211-11
930776e
-3
-2
-1
0
1
2
3
12 51020 50 100 200 500 1k 2k 5k 10k 20k
Typical random-field response with random incidence corrector
Frequency response satisfies ANSI S 1.4–1984 Type 1
-3
-2
-1
0
1
2
3
12 51020 50 100 200 500 1k 2k 5k 10k 20k
Typical free-field response for 0˚ incidence
without random incidence corrector
Tol.
Tol.
Tol.
Tol.
Frequency response satisfies IEC 651 Type 1
1. July 1993
–30.2
1740259
30.9
N.G.