EasyManua.ls Logo

BRUEL & KJAER 4188 - Specifications Overview; Ordering Information

BRUEL & KJAER 4188
219 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 — Prepolarized Free-field
1
/
2
" Microphone Type 4189
Specifications Overview
Falcon Range of Microphone Products
Microphone Handbook
Brüel & Kjær
3
28
3.17 Specifications Overview
3.18 Ordering Information
Preamplifier
Type 2669:
1
/
2
" Microphone Preamplifier
Type 2671:
1
/
2
" Microphone Preamplifier
Calibration Equipment
Type 4231: Sound Level Calibrator
Type 4226: Multifunction Acoustic Calibrator
Type 4228: Pistonphone
UA0033: Electrostatic Actuator
OPEN-CIRCUIT SENSITIVITY (250Hz)
*
:
–26dB
±
1.5 dB re 1V/Pa, 50mV/Pa*
POLARIZATION VOLTAGE:
External: 0 V
FREQUENCY RESPONSE*:
0° incidence free-field response:
10 Hz to 8kHz:
±
1dB
6.3Hz to 20kHz:
±
2dB
In accordance with IEC651, Type 1
LOWER LIMITING FREQUENCY (3 dB):
2Hz to 4Hz (vent exposed to sound)
PRESSURE EQUALIZATION VENT:
Rear vented
DIAPHRAGM RESONANCE FREQUENCY:
14kHz, typical (90
°
phase shift)
CAPACITANCE (POLARIZED:
14pF, typical (at 250 Hz)
EQUIVALENT AIR VOLUME (101.3 kPa):
46 mm
3
*
Individually calibrated
CALIBRATOR LOAD VOLUME (250Hz):
260 mm
3
PISTONPHONE TYPE 4228 CORRECTION:
with DP 0776:
0.00 dB
TYPICAL CARTRIDGE THERMAL NOISE:
14.6dB (A) 15.3 dB (Lin.)
UPPER LIMIT OF DYNAMIC RANGE:
3% distortion:
>146 dB SPL
MAXIMUM SOUND PRESSURE LEVEL:
158 dB (peak)
OPERATING TEMPERATURE RANGE:
–30 to +150
°
C (–22 to 302
°
F)
OPERATING HUMIDITY RANGE:
0 to 100 % RH (without condensation)
STORAGE TEMPERATURE:
–30 to +70
°
C (–22 to 158
°
F)
TEMPERATURE COEFFICIENT (250Hz):
0.001 dB/
°
C, typical (for the range 10 to
+50
°
C)
PRESSURE COEFFICIENT (250Hz):
0.010 dB/kPa, typical
INFLUENCE OF HUMIDITY:
<0.1 dB/100 %RH
VIBRATION SENSITIVITY (<1000 Hz):
Typically 62.5dB equivalent SPL for 1m/s
2
axial
acceleration
MAGNETIC FIELD SENSITIVITY:
Typically 6dB SPL for 80 A/m, 50Hz field
ESTIMATED LONG-TERM STABILITY:
> 1000 years/dB (dry air at 20
°
C)
> 2 hours/dB (dry air at 150
°
C)
> 40 years/dB (air at 20
°
C, 90% RH)
> 1 year/dB (air at 50
°
C, 90% RH)
DIMENSIONS:
Diameter
: 13.2 mm (0.52 in) (with grid)
12.7 mm (0.50 in) (without grid)
Height
: 17.6 mm (0.68 in) (with grid)
16.3 mm (0.64 in) (without grid)
Thread for preamplifier mounting
: 11.7mm –
60 UNS
The data above are valid at 23
°
C, 101.3kPa and
50%RH, unless otherwise specified.

Table of Contents

Related product manuals