4.2.2 Control of special tools :
-Manufacturer control: CCLD Power Supply WB-1372 and WB-1453 and Calibration Exciter Type 4294 must be sent to
Brüel & Kjaer every two (2) years for certification of their specifications.
-On site Control of the CCLD power Supply WB1372 and WB1453:
- Press the switch in battery position. Read the level indicated on the meter: the minimum is: 23 V. If level is under 23 V ,
change the three 9V batteries inside the box.
- Connect the resistance WB-3579 to the CCLD power supply and press the switch in bias position :
o If level indicator is under 11 V , the cable is in short-circuit
o If level indicator is above 17 V , the cable is broken or open circuit