EasyManua.ls Logo

BRUEL & KJAER Mediator 2238 - Introduction to Measurement Setup; Measurement Set-up Menu Navigation; Setting Frequency Weightings

BRUEL & KJAER Mediator 2238
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 — Checking Procedure
Checking Procedure
2238 Mediator Integrating Sound Level Meter
Service Manual
Brüel&Kjær
16
Table 2.6 RMS and Peak Detector Amplitude Linearity
2.3.7 Noise
Action:
Remove the JJ2614 ½² Microphone Adapter.
Mount the WA0302 Microphone Simulator with the short circuit JP0169.
Set range: Press , then until you get the 10-90dB range, then to OK.
Set the Bandwidth/Time Weighting to Broadband/Slow: Press , select
Weightings, then until you get BB/S. Press and to set Frequency
Weighting as given in Table 2.7.
Warning: The display reading for the different Frequency Weightings must be as given in
Table 2.7.
Input
(gen.
output)
Display
dB SPL
Actual
Reading
Display
dB Peak
Actual
Reading
Ref. +36dB 130.0± 0.5 133.0± 0.5
Ref. +26dB 120.0± 0.5 123.0± 0.5
Ref. +16dB 110.0± 0.5 113.0± 0.5
Ref. +6dB 100.0± 0.3 103.0± 0.5
Ref. 94.0 97.0
Ref. 4dB
90.0± 0.3 93.0± 0.5
Ref.
14dB
80.0± 0.5 83.0± 0.5
Ref.
24dB
70.0± 0.5
73.0± 0.7
Ref. 34dB
60.0± 0.5
63.0± 2.0
Ref. 41dB
53.0± 0.5
N/A
Note: N/A = Not Applicable