© Brüel & Kjaer Vibro S1079330.002 / V04 Page 75 of 188
Technical alterations reserved!
Brüel & Kjaer Vibro Instruction
VC-8000 MPS
Content
EN
10.4 Basic Rack Configuration
Follow these basic steps when starting a new
configuration:
1. Start MPS Setup Software
2. Set your default units
(File, Default Units).
3. Select File, New.
4. Select the Modules Tab.
5. (Off-Line Configuration) Select the module type
(for each slot) in the Type column. Make sure to
select the correct slot. If you configure the
module in the wrong slot you will have to start
over.
(On-Line Configuration) If you are connected to
the rack, select GET and the module types will
populate for you.
6. Select Slot 2 (SAM), open the Properties Pane,
and enter the rack name. Typically, this will be
the Machine Train name.
7. Select the Channels Tab.
8. Select the Channel Type for each channel (by
default all channels are Radial Vibration).
Remember Phase Trigger channels can only be
assigned to Channel 4 (Slot 4-9).
9. Enter Channel Tag, Asset 1 (Group), and Asset
2 (Sub-Group). Asset 1 is typically the machine
(i.e. Motor or Pump). Asset 2 is typically the
bearing (i.e. Brg 1 or Brg 2).
10. Save the configuration file (and or SEND to the
rack).
These steps give you a basic layout to your
configuration, and you are now ready to enter the other
configuration parameters.
Note: Changes to the selected Module, or Channel
Type will erase your inputs and return the channel
properties to their default values.
Figure 10-10) Select Module Type
Figure 10-11) Enter Rack Name (identifier)
Figure 10-12) Select Channel Type
Figure 10-13) Enter Channel and Asset Names