BRUEL & KJAER VIBROCONTROL 8000 - 12.2.1 Module Description; 12.2.2 Channel Names

BRUEL & KJAER VIBROCONTROL 8000
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Brüel & Kjaer Vibro S1079330.002 / V04 Page 89 of 188
Technical alterations reserved!
Brüel & Kjaer Vibro Instruction
VC-8000 MPS
Content
EN
12.2.1 Module Description
The module descriptions will be shown in the Maintenance display on the Rack tab of the
Maintenance display. The SAM description should be 8 characters or less. The UMM/TMM
description can be 18 characters.
Figure 12-7) SAM, UMM, TMM description field
12.2.2 Channel Names
Channel names are used on all the display screens to identify the channels. A descriptive channel
name usually works best. A few examples are shown below.
Name (Examples) Comment
VXE-932
VYE-932
VXE-931
VYE-931
Not enough information.
If this is a small installation with a minimal number of points
this could be acceptable.
VXE-932 (T-NDE)
VYE-932 (T-NDE)
VXE-931 (T-DE)
VYE-931 (T-DE)
Good.
The abbreviations used are typical for the United States. T =
Turbine, NDE = Non-Drive End, DE = Driven End.
Turb NDE X (VXE-932)
Turb NDE Y (VYE-932)
Turb DE X (VXE-931)
Turb DE Y (VYE-931)
Good.
For CMS systems this may work better as it allows a more
natural grouping when the channels are sorted
alphabetically.
Motor Lower Bearing Vib X
Motor Lower Bearing Vib Y
Pump NDE Vib X
Pump NDE Vib Y
Good.
These names are long, but there is no possibility of
confusion here.
Gearbox H/S Horz NE-784-22A
C-1A Vib 1_VE-757-74A
H/S Phase Trig NEN-3
Comp OB Horz NE-16A
Hard to read.
These names are long, but there is no possibility of
confusion here.
Use parenthesis to improve readability. For example:
Gearbox H/S Horz (NE-784-22A)
Table 12-1) Channel (tag) name examples