9.5 Balancer
Parameter
Setting para-
meter
Explanation of para-
meter
Plane Setup
Single plane
(A=CH1); Two
plane (A=CH1 &
B=CH2); Two
plane
(A&B=CH1)
Selection of planes for
balancing. Two-plane
balancing can also be
performed sequenti-
ally with a sensor.
Sensor
B&K Vibro sen-
sors; Variable
Sensor type
Sensor Units e.g. g Sensor unit
Sensitivity
e.g. 100 (mV/
Sensor Units)
Sensitivity per sensor
unit.
CCS Power On; Off
Constant Current Sup-
ply. Depends on selec-
tion in Sensor.
Units
g; m/s
2
; mm/s ;
m
Display unit
Detection rms; pc; ppc Signal detection.
Weight Units g; kg; %
Units of weights
applied or removed.
Length Units mm; cm; m; EU Units of length
Trigger
Automatic;
Manual; Visual
(the trigger level
can be visually
displayed and
helps you deter-
mine a trigger
threshold)
Type of triggering
Trigger Slope -ve; +ve
Trigger slope - falling
edge (minus) rising
edge (plus).